Marposs HORIZON · Interferometric Gauge for Multi-layer Materials
A next-generation controller that measures the thickness and distance of wafers and thin films without contact by interferometry. It covers thickness and warpage of silicon, SiC and GaN wafers through in-process measurement during back-grinding.
One controller measures thickness and distance for transparent, translucent and opaque materials, reflective surfaces and advanced semiconductor coatings.
30 nm axial resolution and 0.5–2 µm accuracy depending on model for stable measurement of thin films, wafers and coatings.
Connect multiple units in MASTER-SLAVE for synchronized multi-point measurement, and measure moving targets with encoder synchronization.
Supports Ethernet, USB, RS422, analog output, Anybus (EtherCAT · PROFINET, etc.) and SDK (C++ · C · C#).
Electronic and electromechanical component materials are often built from multiple layers with different properties, some opaque to white light or only a few µm thick. HORIZON measures these thin layers without touching fragile parts. Choose a light source model suited to the layer thickness and material, and sensors with different working distances and sizes for your environment.
| Working Distance (WD) | 10, 28, 50, 100 mm (IP68 optical pen 1.6 mm, usable in DI water) |
|---|---|
| Spot Size | 18 – 30 µm |
| Lateral Resolution | 9 – 15 µm |
| Surface Angle | 90° ± 2° |
| Optical Fiber | 3–4 m, bend radius 30 mm, armored fiber for indoor use |
| Protection · Weight | IP40 · 80 g (excluding cable) |
| Model | S1 / 90S1 | S2 / 90S2 | T1 / 90T1 | W1 | L2 |
|---|---|---|---|---|---|
| Light Source | SLED | SLED | SLED | LED | LED |
| Wavelength | 1310 nm | 1310 nm | 1020 nm | 350–700 nm | 750 nm |
| Measuring Range* | 37–1850 µm | 74–3700 µm | 15–850 µm | 2.25–225 µm | 60–3000 µm |
| Accuracy | ≤ 1 µm | ≤ 2 µm | ≤ 1 µm | ≤ 0.5 µm | ≤ 1 µm |
| Sampling Rate | 2,000 / 90,000 Hz | 2,000 / 90,000 Hz | 2,000 / 90,000 Hz | 2,000 Hz | 2,000 Hz |
* Measuring ranges are based on refractive index n = 1; for other materials, divide by the material refractive index. Ranges may be limited when surface roughness Ra > 0.1. The 90 series (90S1 · 90S2 · 90T1) are high-speed models with a 90,000 Hz sampling rate.
| Principle | Interferometric |
|---|---|
| Measurement Items | Thickness (incl. multi-layer) · distance |
| Axial Resolution | 30 nm |
| Channels | 1 (MASTER-SLAVE multi-point expansion) |
| Encoder Input | Digital 3 (TTL/HTL) |
| Analog Output | 2 (0–10 V DC) |
| Communication | USB · Ethernet · RS422, Anybus slot |
| Software | MIC Tool · SDK (C++ · C · C#) · Quick SPC |
| Power · Consumption | 12–24 V DC · 30 W |
| Protection Class | IP40 |
| Dimensions · Weight | 239 × 157.5 × 131.15 mm · 2.8 kg |
Wafer thickness results measured with HORIZON.
Marposs HORIZON · Interferometric Gauge for Multi-layer Materials
A next-generation controller that measures the thickness and distance of wafers and thin films without contact by interferometry. It covers thickness and warpage of silicon, SiC and GaN wafers through in-process measurement during back-grinding.
One controller measures thickness and distance for transparent, translucent and opaque materials, reflective surfaces and advanced semiconductor coatings.
30 nm axial resolution and 0.5–2 µm accuracy depending on model for stable measurement of thin films, wafers and coatings.
Connect multiple units in MASTER-SLAVE for synchronized multi-point measurement, and measure moving targets with encoder synchronization.
Supports Ethernet, USB, RS422, analog output, Anybus (EtherCAT · PROFINET, etc.) and SDK (C++ · C · C#).
Electronic and electromechanical component materials are often built from multiple layers with different properties, some opaque to white light or only a few µm thick. HORIZON measures these thin layers without touching fragile parts. Choose a light source model suited to the layer thickness and material, and sensors with different working distances and sizes for your environment.
| Working Distance (WD) | 10, 28, 50, 100 mm (IP68 optical pen 1.6 mm, usable in DI water) |
|---|---|
| Spot Size | 18 – 30 µm |
| Lateral Resolution | 9 – 15 µm |
| Surface Angle | 90° ± 2° |
| Optical Fiber | 3–4 m, bend radius 30 mm, armored fiber for indoor use |
| Protection · Weight | IP40 · 80 g (excluding cable) |
| Model | {{ c }} |
|---|---|
| {{ r.k }} | {{ v }} |
* Measuring ranges are based on refractive index n = 1; for other materials, divide by the material refractive index. Ranges may be limited when surface roughness Ra > 0.1. The 90 series (90S1 · 90S2 · 90T1) are high-speed models with a 90,000 Hz sampling rate.
| {{ s.k }} | {{ s.v }} |
|---|