KOR Request a Quote
Desk Top Type Resistivity Meter · 4-Point Probe

Sheet Resistance Meter

A desktop meter that measures sheet resistance, resistivity and resistance of thin films by 4-point probe contact. Single or Dual measurement can be selected, and Ω · Ω/sq · Ω·cm units are converted instantly.

1 mΩ – 1 GΩSheet resistance range (/sq)
± 0.5%Accuracy (RSD-1G)
1 sMeasurement Time
RSD-1G desktop sheet resistance meter
RSD-40k desktop sheet resistance meter

Desk Top Type Resistivity Meter

RSD-1G
Desk Top Type · High Precision

RSD-1G

Single · Dual · Mono · Bulk measurement modes
Instant conversion of Ω, S/cm, Ω/sq, Ω·cm units
Bulk mode for bulk solar cell materials
Dimensions 350 (W) × 240 (L) × 140 (H) mm
Measurement Specifications
Measurement Method4-point probe, contact type
ConfigurationSingle or Dual (selectable)
Measuring RangeSheet resistance 1 mΩ/sq – 1 GΩ/sq Resistivity 1 µΩ·cm – 10 MΩ·cm Resistance 1 mΩ – 250 MΩ
Accuracy± 0.5% (KRISS standard wafer, 23 °C)
Measurement TimeRange search up to 6 s, then 1 s
Utility
Current Source10 nA – 100 mA
PowerAC 100 – 240 V, 50 / 60 Hz
Operating Temperature23 °C ± 5 °C
Relative Humidity30 – 70%
RSD-40k
Desk Top Type · Compact

RSD-40k

Single · Dual · Bulk measurement modes
Instant conversion of Ω, Ω/sq, Ω·cm units
Customizable probe load · tip radius · spacing
Dimensions 145 (W) × 170 (L) × 60 (H) mm
Can be paired with the HS8-um thickness meter
Measurement Specifications
Measurement Method4-point probe, contact type
ConfigurationSingle or Dual (selectable)
Measuring Range1 kΩ/sq – 1 GΩ/sq (1G) 10 mΩ/sq – 40 kΩ/sq (40k)
Accuracy± 1.0% of reading
Measurement TimeRange search up to 6 s, then 1 s
Utility
Current Source10 nA – 100 mA
DVM12 V (DC 4.2 V, 600 mA)
Operating Temperature23 °C ± 5 °C
Relative Humidity30 – 70%

A 4-point probe is required. Probe spacing, tip radius and load can be selected for your samples.

Target

Transparent electrode filmsConductive films for displays and touch panels such as ITO and IZO
Semiconductor WafersSheet resistance · resistivity of wafers and deposited films
Solar CellsBulk mode for bulk solar cell materials
Coatings · Functional FilmsConductive coatings, metal thin films, functional materials

Measurement Examples

Sheet resistance results measured by 4-point probe, by sample type.

ITO transparent electrode sheet resistance example
ITO Transparent ElectrodeTransparent electrode sheet resistance
Metal thin film sheet resistance example
Metal Thin FilmSheet resistance of deposited metal films
Wafer doped layer sheet resistance example
Wafer Doped LayerSheet resistance map of doped wafers
Conductive film sheet resistance example
Conductive FilmConductive coating films

Tell us your sample type and resistance range, and we will propose the right model and probe

T. 042-361-9999 · E. lin-tech@lin-tech.co.kr
Request a Quote · Demo